ANALYZER 90
ANALYZER 90
ANALYZER 90. Reliable and robust semi-automated clinical chemistry 10 mm Flow Cell photometer. Flow-cell with Peltier controlled temperature. Standard configuration with 7 optical filters, wave length 340-400/700 nm. Mono and bichromatic reading systems. Its operating software runs and stores up to 200 tests. Reading system: End-Point, Fixed-Time, and Kinetic. Data processing with different calculation modes.
Download brochure (pdf)
Technical features
Accuracy ± 1% from 0.000 to 1.500 OD
± 2% from 1.500 to 3.000 OD
Stability 0.01 OD in 8 hours
Light source halogen lamp
Optical filters 340, 405, 492, 510, 546, 578, 630 nm
Detector silicon photodiode
Reading path 10 mm
Flow-cell 80 µl
Flow-cell temp room temp. 25/30/37° C ± 0.1° C
Sample Aspiration automated sample aspiration through a
precision peristaltic pump
Sample volume minimum 350 µl
Memory capacity 200 test and curves; the last 300 printable
readings
Reading system mono and bichromatic reading
Fast reading for kinetic and fixed time
Calculation Mode optical density, concentration, endpoint against standard or factor, fixtime against standard or factor, kinetic, non linear multi-standard curve up to 8 points, sample blank, reagent blank Calibration from 2 to 8 points curve
Calendar/Timer associates time with the results; timer function
Keypad 17 buttons
Display 2 lines with 16 columns
Printer 40 columns with thermal paper
Interface RS232 serial port
Warm up Time 15 minutes
Power requirements 230/115 VAC, 50-60 Hz, 110 W
Measurements 42 x 34 x 15 cm
Weight 9 Kg