| ANALYZER 90 |
 |
| ANALYZER 90. Reliable and robust semi-automated
clinical chemistry 10 mm Flow Cell photometer.
Flow-cell with Peltier controlled temperature.
Standard configuration with 7 optical filters,
wave length 340-400/700 nm.
Mono and bichromatic reading systems.
Its operating software runs and stores up to 200
tests. Reading system: End-Point, Fixed-Time,
and Kinetic. Data processing with different
calculation modes.
|
|
|
| Technical features |
± 1% from 0.000 to 1.500 OD
± 2% from 1.500 to 3.000 OD |
| 0.01 OD in 8 hours |
| halogen lamp |
| 340, 405, 492, 510, 546, 578, 630 nm |
| silicon photodiode |
| 10 mm |
| 80 µl |
| temp room temp. 25/30/37° C ± 0.1° C |
automated sample aspiration through a
precision peristaltic pump |
| minimum 350 µl |
200 test and curves; the last 300 printable
readings |
| mono and bichromatic reading |
| for kinetic and fixed time |
| optical density, concentration, endpoint against standard or factor, fixtime against standard or factor, kinetic, non linear multi-standard curve up to 8 points, sample blank, reagent blank Calibration from 2 to 8 points curve |
| associates time with the results; timer function |
| 17 buttons |
| 2 lines with 16 columns |
| 40 columns with thermal paper |
| RS232 serial port |
| 15 minutes |
| 230/115 VAC, 50-60 Hz, 110 W |
| 42 x 34 x 15 cm |
| 9 Kg |
|
|
 |