| AP22 Speedy IF |
 |
| AP22 Speedy IF is a fully automated micro-plate and slide processor for ELISA and IFA methods.
The instrument is able to manage 2 ELISA micro-plates with 8 methods on line or 16 IFA slides, from sample
pre-dilution and reagent/dispensation, incubation and washing cycles to the Final reading and data processing.
Thanks to its technical knowledge DAS has been able to merge both technologies, ELISA and IFA assays on the
innovative, powerful and reliable AP22 Speedy IF system. |
|
|
| Technical features |
| 60 / 80 – Ø 12 mm sample tubes or 40 – Ø 16 mm sample tubes |
| 92 positions |
| 92 or 200 positions |
| modular racks to hold calibrators, controls and reagents for 8 methods on line |
| 5 external reservoirs: 4 with level sensor (BUFFER 1, 2, wash and waste), 1 without level sensor (slide PBS) |
| 2 standard plates (12 x 8 well format) with programmable incubation from room temperature to 45 °C |
2 probes one for dispensation and dilution, the other for slide washing
|
|
two 8 nozzle manifolds one for dispensation and the other for aspiration
|
8 independent photo-metric channels with mono and bichromatic reading capability
|
| 2 precision filters (450 and 630 nm); 2 more on request (4 filters max, 400-700nm) |
| reading from 0.000 to 3.000 OD, inaccuracy ± 1% from 0.000 to 1.500 OD; ± 2% from 1.500 to 3.000 OD |
| from 8 µl to 1000 µl by steps of 1 µl, inaccuracy ± 1% with 100 µl |
IFA TECHNICAL FEATURES
| 20s with 1ml of washing solution |
| 9s with 1ml of washing solution |
| 42s for a 12 well slide for 20 µl each (including washing with 1 ml buffer) |
CV < 2% at 100 µl
|
CV < 2% at 100 µl
|
|
1 tray with 16 positions for IFA slides; 1 reagent tray for 12 controls, 4 conjugates (3 in case of Blue Evans dye), 1 washing solution for the probe and 1 diluent for samples
|
| sensitivity able to detect 150 µl |
| 2 probes flux or drop washing system for the same well, one to dispense and the other for aspirate |
OPERATING FUNCTIONS
The instrument can manage 8 methods (IFA 2) with sample diluent and a washing Buffer (Blue Evans dye optional) in the same run
• Slides mask configuration
• Wells shape and size configuration
• Total number of wells per slide
• Number of well lines, max 2 per slide
• Coordinates of the first and last well of each line
• Slide thickness
• Controls and samples in double dispensation |
• Distance between probes by 1/10 mm steps
• Distance between the dispensation probe tip and slide surface by 1/10 mm steps
• Distance between the aspiration probe tip and slide surface by 1/10 mm steps
• Washing cycle duration from 1 to 10s (flow washing)
• Number of washing cycles from 1 or 2
• Dispensation flow from 100 µl/s to 900 µl/s on the slide (flow washing)
• Drops numbers (drop by drop)
|
• Possibility to define incubation time and dispensation volume for samples and conjugates
• “Blue Evans dye” settings
• Possibility to set different incubation times for each method on the same run
• Programmable dilutions: 10 for ELISA, 12 for IFA
• Serial dilution onto 2 predilution cuvettes
• The last session work list printing
• Work list traceability
• Printing of trays configuration: sample, predilution, reagent and slides trays
• Assay set up printing
• Slide set up printing
• User-friendly software program, with on-line help
• Memorization of the programmed settings
• Host Computer Work list set up
• Possibility of bypassing the dilution phase
• Possibility of setting the slide thickness for each method on the same run
• Display and printing of the undetected liquid |
LIS interface capability, work list set up from Host computer. Capability of connecting more instruments to a single PC
|
| 230/115 VAC, 50-60 Hz, 350 W |
cm 62 x 82 x 72h
|
| 80,5 Kg |
| sample Bar Code Reader |
|
|
 |